BENDER MEDICAL | BACK |
A-ISOMETER® isoMED427P
The A-ISOMETER® isoMED427P monitors the
insulation resistance of unearthed AC circuits (medical "IT systems").
At the same time, the load current and temperature of the IT system
transformer is monitored.
Insulation monitoring device with
integrated load and temperature monitoring and locating current injector
for insulation fault location systems for medical IT systems in
accordance with IEC 60364-7-710, IEC 61557-8, IEC 61557-9 and DIN VDE
0100-710
In combination with EDS series insulation
fault locators and the appropriate measuring current transformers, the
isoMED427P is designed to set up the respective equipment for insulation
fault location |
|
EDS151
The insulation fault locator EDS151 in
conjunction with the A-ISOMETER® iso-MED427P, the automatic transfer
switching device ATICS or the locating current injector PGH474, are
designed for insulation fault location in unearthed power supplies (IT
systems). The locating current pulse generated by the iso-MED427P, ATICS
or PGH are detected using the integrated measuring current transformers
and evaluated by the insulation fault locators. The integration of six
measuring current transformers in an EDS151 permits all current-carrying
conductors of an outgoing line to be routed through. The response time
for an alarm message inclusively indication on the respective display
device is max. 8 s (e.g. MK2430).
A total of 88 EDS151 devices can be connected via
an RS-485 interface (BMS protocol). Hence, up to 528 circuits can be
monitored. Activities on the BMS bus are indicated by an alarm LED. |
|